Test ID BLOD0220 Zinc, Serum

Reporting Name

Zinc, S

Useful For

Detecting zinc deficiency

Performing Laboratory

Mayo Medical Laboratories in Rochester

Specimen Type

Serum


Specimen Required


Collection Container/Tube: 6-mL Plain, royal blue-top Vacutainer plastic trace element blood collection tube (Supply T184)

Submission Container/Tube: 7-mL Mayo metal-free, screw-capped, polypropylene vial (Supply T173)

Specimen Volume: 0.8 mL

Collection Instructions:

1. Allow the specimen to clot for 30 minutes; then centrifuge the specimen to separate serum from the cellular fraction. Serum must be removed from cellular fraction within 4 hours of specimen collection. Avoid hemolysis.

2. Remove the stopper. Carefully pour specimen into a Mayo metal-free, polypropylene vial, avoiding transfer of the cellular components of blood. Do not insert a pipet into the serum to accomplish transfer, and do not ream the specimen with a wooden stick to assist with serum transfer.

3. See Trace Metals Analysis Specimen Collection and Transport in Special Instructions for complete instructions.

Additional Information:

1. High concentrations of gadolinium, iodine, and barium are known to interfere with most metals tests. If gadolinium-, iodine-, or barium-containing contrast media has been administered, a specimen should not be collected for 96 hours.

2. If ordering the trace element blood collection tube from BD, order catalog #368380.


Specimen Minimum Volume

0.2 mL

Specimen Stability Information

Specimen Type Temperature Time
Serum Refrigerated (preferred) 14 days
  Ambient  14 days
  Frozen  14 days

Reference Values

0-10 years: 0.60-1.20 mcg/mL

≥11 years: 0.66-1.10 mcg/mL

Day(s) and Time(s) Performed

Monday; 2 p.m.; Tuesday through Friday; 5 p.m.; Saturday; 2 p.m.

Test Classification

This test was developed and its performance characteristics determined by Mayo Clinic in a manner consistent with CLIA requirements. This test has not been cleared or approved by the U.S. Food and Drug Administration.

CPT Code Information

84630

LOINC Code Information

Test ID Test Order Name Order LOINC Value
ZNS Zinc, S 5763-8

 

Result ID Test Result Name Result LOINC Value
8620 Zinc, S 5763-8

Analytic Time

1 day

Reject Due To

Hemolysis

Mild reject; Gross reject

Lipemia

Mild OK; Gross OK

Icterus

Mild OK; Gross reject

Other

NA

Method Name

Dynamic Reaction Cell-Inductively Coupled Plasma-Mass Spectrometry (DRC-ICP-MS)

Forms

If not ordering electronically, complete, print, and send a General Request Form (T239) with the specimen (http://www.mayomedicallaboratories.com/it-mmfiles/general-request-form.pdf).